Review Article

Complexity and Perplexity of Pulsed Field Ablation: An Engineering Perspective

Register or Login to View PDF Permissions
Permissions× For commercial reprint enquiries please contact Springer Healthcare: ReprintsWarehouse@springernature.com.

For permissions and non-commercial reprint enquiries, please visit Copyright.com to start a request.

For author reprints, please email rob.barclay@radcliffe-group.com.
Information image
Average (ratings)
No ratings
Your rating

Abstract

Pulsed field ablation (PFA), recently introduced as a non-thermal and selective method for cardiac ablation, was associated with great promise, hope and expectation, but also raised some concerns and left some questions unanswered, in particular with respect to waveform. To better understand the challenges associated with the design and development of safe and efficient PFA systems, the underlying mechanism of electroporation at the membrane, cellular and tissue levels is described. The three interdependent components of each system, that is, the waveform, the catheter and the generator, are then addressed. The effect of the different waveform parameters on treatment outcomes is reviewed, and the consequences of a potential mismatch of the three components in the development of a safe and efficient PFA system are highlighted.

Received:

Accepted:

Published online:

Disclosure: BK is supported by the Slovenian Research Agency (ARIS) in research projects No. J2–50068 and P2–0249 and has received consulting fees from Medtronic and travel support from Mediasphere. MR is supported by the Slovenian Research Agency (ARIS) in research project No. J2–50068 and has a 70% share in mPOR d.o.o. DM is supported by the Slovenian Research Agency (ARIS) in research project No. P2–0249, has received grants from Medtronic, royalties from Springer Nature and IGEA, consulting fees from Medtronic and Inomagen Therapeutics, honoraria from Medtronic, Boston Scientific and Field Medical, and accommodation support from Field Medical, and has a 10% share in mPOR d.o.o.

Correspondence: Damijan Miklavčič, Faculty of Electrical Engineering, University of Ljubljana, Tržaška cesta 25, 1000 Ljubljana, Slovenia.
E: damijan.miklavcic@fe.uni-lj.si

Copyright:

© The Author(s). This work is open access and is licensed under CC-BY-NC 4.0. Users may copy, redistribute and make derivative works for non-commercial purposes, provided the original work is cited correctly.

Pulsed field ablation (PFA), recently introduced as a non-thermal and selective method for cardiac ablation, is associated with great promise, hope and expectation, but also raises some concerns.1–3 Unfortunately, from a scientific and engineering perspective, PFA is associated with a poorly defined design and parameter space due to the nature of the treatment, which includes load variability and requires multi-parameter optimisation with several potentially conflicting constraints. The device, that is, the waveform, the catheter and the pulse generator, forms the trinity of PFA. They must be developed together and function as a whole that is greater than the sum of its parts. An ever-increasing number of newly developed pulse generators and catheters with different waveforms raises important questions. Are they comparable and does a certain combination have specific side effects? Are these specific or the same for all systems? Do we need to ask the same questions and conduct the same studies for each new PFA system?

To better understand the challenges of developing a PFA system, we will first describe the phenomenon of electroporation that underlies PFA at the membrane, cellular and tissue levels. We will then look at the waveform, the catheter and the pulse generator, which must be considered and developed as a unit to be fully functional. Even small changes in one of the three components can cause the whole system to fail or at least operate suboptimally. Only by understanding these aspects can we fully assess the challenges and recognise how narrow the path to success can be.

Basic Description and Understanding of Electroporation

Cell Membrane: Increased Conductivity and Cell Depolarisation

The cell membrane separates the inside of the cells from the outside. It has a very selective permeability for ions and molecules, which enables the cell to survive even in a sometimes somewhat unstable environment. In excitable cells, ion channels and pumps ensure that the cells can generate and transmit action potential. The cell membrane can be regarded as a capacitor from an electrical point of view and thus represents a barrier for electrical current (at low frequencies). During and after electroporation, the conductivity of the cell membrane is greatly increased and the membrane remains permeable to ions and other molecules for up to several minutes after treatment.4,5 Even though this persistent increase in permeability for ions (e.g. Na, Cl, Ca, K, etc.) is smaller than during the pulse delivery itself, it is sufficient to cause and maintain cell depolarisation, which can be transient (triggering action potential) or can result in sustained depolarisation.6–8 Increased transient membrane permeability results in the stunning of excitable cells (rendering them unexcitable or causing conduction block).9 This depolarisation can result in immediate disappearance of local electrograms and transient phrenic nerve paresis.10–13 Given that the membrane damage caused by electroporation is followed by membrane repair, this can lead to cells regaining the ability to react to an electrical stimulus within a few minutes, depolarising and conducting an action potential.14,15 For a given pulse duration and number of pulses, the most important parameter that determines the level and intensity of electroporation is the amplitude of the electric field to which the cell is exposed.

Electroporation at the Cellular Level: Reversibility

As a consequence of cell membrane electroporation and increased membrane permeability, there are several downstream effects, including changes in gene regulation and protein expression.5 Historically, electroporation was separated into reversible and irreversible, with the only clear determination between the two represented by whether the cells survive the treatment or later die via one of the cell death pathways.16 Reversible electroporation is typically associated with applications such as drug and gene delivery, in which transient permeabilisation of the cell membrane enables therapeutic agents to enter the cell before the membrane permeability returns to normal physiological conditions and the cell therefore recovers.17–20 Irreversible electroporation, in contrast, leads to changes in the membrane or sustained disruption, which leads to cell death via diverse cell death pathways.21–23

Membrane resealing alone, however, does not guarantee cell survival. Cell death is a dynamic process, and different pathways of cell death can occur in the same lesion at different times, locations and distances from the catheter.24,25 The electric field closest to the catheter and electrodes is the highest and then drops rapidly with distance from the catheter.26 It is therefore plausible to speculate that necrotic or pyroptotic cell death predominates in the lesion core (where the field is strongest), while apoptotic mechanisms may be more prevalent at the lesion periphery.

Tissue Level: Electric Field Distribution and Cell–Cell Interactions

To achieve therapeutic electroporation in tissue, an electric field has to be established in the tissue, which is usually achieved by bringing electrodes in contact with the tissue. In cardiac electrophysiology, this is most commonly achieved in a minimally invasive way using a catheter approach (Figures 1A and 1B). The cells of the tissue are organised and embedded in the extracellular matrix. Several different cells coexist in the same volume of tissue, nerves pass through, and vessels bring oxygen (and nutrient-rich blood) to every cell in the body. Electroporation occurs at the membrane cell level, as described above, and all cells can be electroporated, including (but not limited to) cardiomyocytes, fibroblasts, neurons, endothelial cells and erythrocytes (Figure 1C). In addition to the effects on individual cells described in the previous sections, cell–cell interactions are at least transiently disrupted, leading to leaky vessels, which results in oedema formation.27 All of these effects occur simultaneously due to high-voltage pulse delivery, but they have different dynamics of resealing and recovery.

Figure 1: Pulsed Field Ablation at Different Scales

Article image

The membrane conductivity increase due to electroporation also leads to an increase in tissue admittance (i.e. a decrease in tissue impedance). This means that the electrical load is changing during the delivery of pulses in a non-linear fashion.28,29 Based on previous in vivo studies on gene transfer and drug delivery, it is well established that pulsed electric fields transiently reduce tissue perfusion and increase vascular permeability, including temporary disruption of the blood–brain barrier.30–32 These changes lead to a reduction of tissue cooling due to diminished or absent capillary blood flow, and promote oedema formation. The resulting oedema further decreases tissue impedance following pulse delivery and contributes to the early stages of wound healing and tissue repair.33,34 Additionally, reduced perfusion and elevated interstitial fluid pressure caused by oedema may impair the contractile function of cardiomyocytes.

While PFA was initially described as more selective for cardiac tissue based on in vitro data, this is not mirrored by in vivo studies and their findings of a lethal electric field.35–37 An interesting observation is that PFA can ablate through scarred tissue.38 Scarred tissue, which is mostly acellular, has a significantly higher conductivity than healthy myocardium.39,40 Scarred tissue seems to present a lower barrier for the electric field, however, with the increase of conductivity due to electroporation, the electroporated myocardium becomes similarly conductive as the scarred tissue, which results in the ability of PFA to ablate through the scar.26

It is important to note that electroporation is a physical mechanism of cell membrane disruption that can affect all cell types. For PFA, the most relevant cells are of course cardiomyocytes, but other cells such as erythrocytes, neurons, cardiac fibroblasts and cells of the cardiac conduction system can also be electroporated. All of the electroporation effects on the cells described above can also be observed in these cells, leading, for example, to haemolysis, spasm of the cardiac arteries, phrenic nerve paresis, and disruption of the cardiac conduction system.41–50 It is not yet clear to what extent these effects are reversible, and what affects the extent of injury and rate of recovery of their normal function.

While the electric field decreases rapidly with distance from the catheter surface, it is important to consider that the electric field at a given point in the tissue (Figure 1B) depends on the geometry of the catheter, the tissue and its electrical properties, and the configuration of the return electrode. If all of these factors are kept constant, the electric field depends on the voltage.

The Trinity of PFA: Waveform, Catheter and the Pulse Generator

Pulsed field ablation depends on the successful delivery of a sufficiently strong electric field in the target tissue. As shown in Figure 2, the waveform, catheter and pulse generator must function together as a whole, greater than the sum of its parts. The waveform must effectively irreversibly electroporate the targeted cells (i.e. cardiomyocytes) of the arrhythmogenic substrate and avoid, as much as possible, heating and bubble formation, minimise neuromuscular capture (pain and muscle contraction) and reduce or minimise stunning and/or reversible electroporation.51 For atria and specifically for pulmonary vein isolation (PVI), 2–5-mm-deep lesions are sufficient to create transmural lesions and effectively isolate the pulmonary veins. However, for targeting ventricular substrate, this is not sufficient because greater depth is required. In the following sections we take a look at the three components of the trinity.

Figure 2: The Pulsed Field Ablation Trinity

Article image

The Waveform

Early electroporation research was performed using monophasic 100 μs pulses.52 Nanosecond pulses and sub-microsecond pulses were then intensively researched, because they looked promising for causing interesting biological effects.53 Later, Arena et al. suggested using biphasic short pulses with the intention of reducing the contrast in tissue conductivity and neuromuscular stimulation.54 Several first-in-human studies of PFA were performed using monophasic pulses, but those quickly switched to biphasic pulses.55 A biphasic waveform has many parameters, all of which have the possibility of affecting the treatment outcomes.56–58 Figure 3A shows a compact and complete set of waveform parameters. A single treatment waveform can be composed of several trains. Each individual train can be composed of a single or several bursts of pulses. A single burst can contain one or several biphasic pulses. The total duration of a treatment then depends on the number of trains, the duration of each train and the delay between trains.

Figure 3B shows how the different parameters of the pulses affect treatment. An increase in pulse amplitude greatly increases irreversible electroporation and heating, and causes a small increase in all other outputs.10,24,25,37,59,60 Increasing pulse width causes a large increase in electrochemical reactions, pain, neuromuscular capture and arrhythmogenicity.61–63 Increasing the number of pulses causes a large increase in electrochemical reactions, by increasing the total amount of charge delivered, and a small increase in all other parameters.

Figure 3: Definition of Waveform Parameters and Their Effects

Article image

The shape of the pulses has a very important effect: we know that longer monophasic pulses are more efficient in electroporating cells and that monophasic pulses delivered at a relatively low repetition rate should be the preferred choice.25,37 However, monophasic pulses cause electrochemical reactions and severe neuromuscular capture and pain.22,55,64–66 Biphasic pulses dramatically reduce electrochemical reactions, but also decrease all other parameters except heating, which is unaffected by pulse shape, and depends only on the total energy and rate of pulse delivery.37,62–64,67 Increasing interpulse delay reduces heating slightly, but also greatly increases pain, neuromuscular capture and arrhythmogenicity.62,63,67 Intertrain delay has an effect on the heating: a long pause between pulse trains enables cooldown of the tissue by blood (and to a smaller extent by catheter irrigation).68

The Catheter

The catheter must be manoeuvrable, should be introduced through a small-diameter sheath, able to conduct a high-voltage electric signal from the generator to electrodes, and sustain the high voltage in the limited space of the connecting cables and catheter. The catheter design can be unipolar, meaning that the pulses are delivered between the active electrodes on the catheter and a grounding electrode on the surface of the patient; or bipolar, indicating that the pulses are delivered between electrodes on the same catheter. The electrodes on the catheter have to provide an effective distribution of electric field in the tissue, and minimise local heating and stray fields that are generated in the blood pool (risk of haemolysis) and also in tissue distant from the target area causing nerve and muscle stimulation (neuromuscular capture). In this respect unipolar PFA delivery should reduce haemolysis but also increase neuromuscular capture compared to bipolar delivery. Regardless of the specific catheter design, the constrained shaft (internal space) can cause the delivered waveform to the tissue to deviate from the waveform at the output of the generator resulting in reduced voltage and altered pulse shape (Figure 4A–C).

Most commercial radiofrequency (RF) or PFA cardiac catheters have catheter cabling that acts as a low-pass filter with a frequency cut-off around 2 MHz. As a result, pulse rise and fall times are prolonged by approximately 200 ns. Consequently, square wave pulses longer than 1 µs undergo minimal distortion, while a biphasic pulse requires at least a 200 ns interphase delay (i.e. the pause between the positive and negative phases) to maintain charge balance. However, in such a system, a 200 ns pulse loses 50% of its power and takes on a triangular shape (Figure 4C). Therefore, to enable efficient nanosecond pulse delivery, catheter cabling and active delivery electrodes must be optimised accordingly.

In contrast to RF ablation, PFA does not depend on conductive diffusion transfer to achieve ablative effect in tissue. In fact, blood is a tissue with some of the highest conductivity in the human body.69,70 This means that electric fields will spread through blood in a generally similar way as through the myocardium. During the pulse delivery itself conductivity of tissue increases. Given the conductivity increase factor for cardiac tissue reported in the literature, the conductivity of myocardial tissue affected by electroporation becomes similar to that of blood.37 Consequently, when the catheter is not in direct contact with the myocardium, the depth of the resulting lesion is reduced. This reduction in lesion depth is usually at least equal to the distance between the catheter and the myocardium.71 It is important to note that in PFA – in contrast to RF ablation – achieving good contact is more important than contact force for ensuring optimal lesion size.72–78

The Pulse Generator

Like most new medical devices, the main design of the current PFA systems was locked for several years before they were approved for the market. Figure 4D shows a typical design cycle of a medical device.79 This cycle is very long, given that many steps are required to develop such a treatment. At each step of the process, detailed evaluation is performed and some parameters of the system are locked. First, new hardware must be developed, which must then be tested for electrical safety and electromagnetic compatibility. Then the preclinical tests are carried out, followed by clinical trials. During this period it is difficult to change the hardware because it has already passed previous tests, therefore the outcome is often already known at the beginning of the cycle. However, there was a lack of sound knowledge about this treatment when the first decisions were made. Therefore, many of the existing PFA systems were modified at very late stages (Figure 4D).56–58 However, during this development cycle, many new insights have been gained in preclinical and clinical trials and in the widespread adoption of the technology. This can constitute a basis for a new design cycle. Unfortunately, many of these findings cannot be used collectively to improve PFA because the waveforms of PFA systems are not publicly disclosed and shared between systems. Hence, our knowledge is only partially assembled and far from complete.

Figure 4: Development of a Pulsed Field Ablation System: From Concept to Reality

Article image

After a decade of studies, we now know that shorter pulses attenuate unwanted muscle contractions and nerve stimulation, but this also reduces the efficiency of the treatment, which can be compensated for by applying higher voltages.25,54,61,62,80 A compromise between side effects and efficiency is therefore being sought, which has led to the development of generators with high voltage and short pulses. However, the higher the voltage and the shorter the pulses, the more difficult it becomes to develop such pulse generators and, above all, to ensure that the device is safe for the patient and the operator and does not interact with other devices in crowded electrophysiology laboratories.

The development of silicon carbide switches has now made it possible to develop high voltages faster, and in shorter pulses, which attenuates muscle contractions and nerve stimulation.81 However, faster pulse rise times also increase peak leakage currents, peak electromagnetic interference and peak voltages across the reinforced insulation.82 Designers in this area should pay particular attention to this, given that PFA waveforms generate atypical interference. Typical medical devices generate continuous leakage currents and electromagnetic interference. However, PFA systems generate high peak and low RMS (root mean square) leakage currents and quasi-peak electromagnetic interference due to their fast rise and fall times and long delays between pulses and bursts. Such a device can pass the standardised leakage current and electromagnetic compatibility (EMC) tests, given that the medical device standard requires measurements only of low-frequency leakage currents that can lead to cardiac arrest and quasi-peak EMC.83,84 However, this does not mean that the high peak leakage current cannot affect the patient’s untargeted tissue and that high electromagnetic spikes cannot interfere with some of the neighbouring devices.

In addition, transient overvoltages occur across the reinforced insulation in PFA systems during PFA delivery, hence it is not sufficient to base the insulation design on the steady-state operating voltage alone. These transient overvoltages must be considered when designing the reinforced insulation.85 Fortunately, existing standards for medical devices include methods for calculating insulation requirements that take into account such transient conditions. High-voltage pulses with fast rise times, as used in PFA, can lead to electrical discharges and electric arcing. During arcing, the current can increase significantly, which poses a risk to the patient’s health and can lead to damage to the components of the pulse switch. To mitigate these risks, current- and energy-limiting circuitry should be incorporated into the PFA system to prevent arcing and protect critical switching elements.86

Compared with RF ablation signals, PFA signals have significantly higher amplitudes (in the order of kilovolts), while measurement signals used for mapping, temperature and force detection are typically in the millivolt range. This large difference in signal magnitude makes electrical isolation within the confined wiring of cardiac catheters and connectors particularly challenging. In addition, switching between high-voltage and high-current PFA pulses and low-voltage and low-current measurement of intracardiac electrogram (iEGM) signals in switching units presents a further design and engineering challenge.

Another unmet need is periprocedural guidance of PFA. It is impossible to reliably predict durable lesion based on bipolar iEGM signal attenuation and voltage maps due to transient stunning of the cardiomyocytes (i.e. reversible electroporation). Early systems were not (well) integrated into mapping systems for catheter visualisation; and contact assessment was not available. Given that the lack of contact reduces lesion depth, the reliability and durable efficacy were limited. Clinical experience now shows that repositioning and following the protocols as prescribed by manufacturers is essential to achieve PVI.87,88

Pulse generation technology and catheter technology have changed significantly since the initial development of the hardware, and we have much more data on the clinical efficacy and side effects of specific pulse waveforms and electrical geometries. This could mark the beginning of a new cycle of PFA systems, 2.0.

Effects of Mismatched Waveform and Catheter Design

The shape of the electrodes, and the positioning of and spacing between the electrodes on the catheter can be precisely defined and fixed (circular loop catheter) or have a variable geometry (pentaspline and variable loop catheters).1 These influence not only the distribution of the electric field, but also the load – that is, how much current will flow through the wires in the catheter shaft during pulse delivery. The currents can easily be in the range of 10–30 A, which corresponds to a high instantaneous power.

The different sizes of the catheter, the shape (e.g., flower or basket), and vectoring (between splines, bipolar or unipolar) determine the distribution of the electric field, as well as the load — the amount of current the generator must provide. The distribution of the electric field also determines the size and depth of the lesion, but due to the different waveforms used, the same electric field threshold cannot be used to compare different catheters. A catheter (as a load) in contact with the tissue behaves differently to a catheter in the blood pool, that is, in slight or partial contact with the tissue. Furthermore, the form factor of the catheter will also affect lesion size dependence on contact force.

PFA catheters are available in very different shapes and sizes. Although the manufacturers do not provide information on the waveforms, various protocols with preclinical results are described in the literature. To illustrate the effects of mismatched waveforms and catheters, we used numerical modelling to examine all possible combinations of three catheters and three waveforms described in the literature.

The three catheters were a generic decapolar loop catheter, an 8 mm spherical tip catheter, and a custom bipolar balloon catheter.47.89,90 Each catheter was tested with three pulse protocols: a single 6 ms monophasic exponentially decaying defibrillator pulse; 90 × 100 µs monophasic pulses delivered at 1 Hz (i.e. irreversible electroporation; IRE); and 10 trains of a single burst of 333 biphasic pulses with a 3 μs positive pulse width, 0 μs intraphase delay, 3 μs negative pulse width and 0 μs interpulse delay (i.e. high-frequency irreversible electroporation; HFIRE).65,66,90 Pulse amplitudes were adjusted for each configuration to achieve transmural lesion depth (3 mm in the schematic atrium). More details on the modelling are available in the Supplementary Material.

Electrode surface temperatures varied significantly across combinations. Generally, the HFIRE waveform resulted in the highest surface temperature rise, due to the fact that the required amplitude was the highest, and also because the waveform has a 100% duty factor (defined as the total time in the ‘on’ position divided by the total duration of the pulse train). In these configurations the monopolar deliveries resulted in lower electrode surface temperatures (Figure 5A and D). The 8 mm tip catheter resulted in the lowest average electrode surface temperature, however, the total surface area of the electrode was also the largest, and it enables only point-by-point ablation, whereas the other electrode configurations theoretically represent a single-shot approach. The lowest surface temperatures were observed with the long IRE protocol, which has a very low duty factor and a long total duration. It is therefore almost completely mitigated by blood flow cooling (or by diffusion in Figure 5C, in which blood flow is blocked by the balloon). The investigated balloon also had a very high surface temperature in the HFIRE configuration, due to the mismatch between the surface area of the two delivery electrodes: namely, the tip electrode was much smaller than the ring electrode positioned at the PV ostium. Therefore, the tip electrode has a much higher local current density and resultant higher heating.

Figure 5: Time Course of Electrode Surface Temperature

Article image

Conclusion

A well-designed PFA system should be safe and efficient. Currently available systems (i.e. those that are approved or are being developed and tested) were mostly developed for AF treatment, that is, to achieve PVI. PFA offers unique opportunities: for the first time we do not need to compromise on effectiveness in PVI for safety. This should enable us to test hypotheses that are driving the treatment of paroxysmal and persistent AF patients with much better precision than was previously possible. At the same time, given that PFA is at least as effective as RF ablation and cryo-balloon ablation but has superior efficiency, this will enable the treatment of increasing volumes of patients early after their initial diagnosis.91,92

Click here to view Supplementary Material.

Clinical Perspective

  • A successful pulsed field ablation (PFA) procedure depends on the seamless integration of three key system components: the waveform, the catheter and the generator, which must be designed and calibrated to work in harmony for optimal therapeutic effect.
  • A basic understanding of the mechanisms of electroporation is crucial to ensure consistent, safe and effective PFA treatment in different clinical scenarios. Failure to do so may result in treatment inconsistencies and unexpected outcomes.
  • Current PFA workflows are not interchangeable between different devices, meaning that what works for one system should not be adopted for another without adequate research and clinical validation.

References

  1. Chun KJ, Miklavčič D, Vlachos K, et al. State-of-the-art pulsed field ablation for cardiac arrhythmias: ongoing evolution and future perspective. Europace 2024;26:euae134. 
    Crossref | PubMed
  2. Lakkireddy D, Katapadi A, Garg J, et al. NEMESIS-PFA: investigating collateral tissue injury associated with pulsed field ablation. JACC Clin Electrophysiol 2025;11:1747–56. 
    Crossref | PubMed
  3. Sugrue A, Shabtaie S, Tan NY, et al. Considerations regarding safety with pulsed field ablation for atrial fibrillation. Heart Rhythm O2 2024;5:655–61. 
    Crossref | PubMed
  4. Kotnik T, Rems L, Tarek M, Miklavčič D. Membrane electroporation and electropermeabilization: mechanisms and models. Annu Rev Biophys 2019;48:63–91. 
    Crossref | PubMed
  5. Jan V, Jusović M, Miklavčič D. Effects of electroporation on the function of sarco/endoplasmic reticulum Ca2+-ATPase and Na+,K+-ATPase in H9c2 cells. Appl Sci 2024;14:2695. 
    Crossref
  6. Casciola M, Xiao S, Pakhomov AG. Damage-free peripheral nerve stimulation by 12-ns pulsed electric field. Sci Rep 2017;7:10453. 
    Crossref | PubMed
  7. Azarov JE, Semenov I, Casciola M, Pakhomov AG. Excitation of murine cardiac myocytes by nanosecond pulsed electric field. J Cardiovasc Electrophysiol 2019;30:392–401. 
    Crossref | PubMed
  8. Blažič A, Guinard M, Leskovar T, et al. Long-term changes in transmembrane voltage after electroporation are governed by the interplay between nonselective leak current and ion channel activation. Bioelectrochemistry 2025;161:108802. 
    Crossref | PubMed
  9. Rems L, Rainot A, Wiczew D, et al. Cellular excitability and ns-pulsed electric fields: potential involvement of lipid oxidation in the action potential activation. Bioelectrochemistry 2024;155:108588. 
    Crossref | PubMed
  10. Koruth JS, Neuzil P, Kawamura I, et al. Reversible pulsed electrical fields as an in vivo tool to study cardiac electrophysiology: the advent of pulsed field mapping. Circ Arrhythm Electrophysiol 2023;16:e012018. 
    Crossref | PubMed
  11. Cespón-Fernández M, Pannone L, Sieira J, et al. Feasibility of reversible electroporation mapping in human atrial flutter. Heart Rhythm 2025; epub ahead of press. 
    Crossref | PubMed
  12. Franceschi F, Koutbi L, Gitenay E, et al. Electromyographic monitoring for prevention of phrenic nerve palsy in second-generation cryoballoon procedures. Circ Arrhythm Electrophysiol 2015;8:303–7. 
    Crossref | PubMed
  13. Pansera F, Bordignon S, Bologna F, et al. Catheter ablation induced phrenic nerve palsy by pulsed field ablation – completely impossible? A case series. Eur Heart J Case Rep 2022;6:ytac361. 
    Crossref | PubMed
  14. Chaigne S, Sigg DC, Stewart MT, et al. Reversible and irreversible effects of electroporation on contractility and calcium homeostasis in isolated cardiac ventricular myocytes. Circ Arrhythm Electrophysiol 2022;15:e011131. 
    Crossref | PubMed
  15. Hunter DW, Kostecki G, Fish JM, et al. In vitro cell selectivity of reversible and irreversible: electroporation in cardiac tissue. Circ Arrhythm Electrophysiol 2021;14:e008817. 
    Crossref | PubMed
  16. Yarmush ML, Golberg A, Serša G, et al. Electroporation-based technologies for medicine: principles, applications, and challenges. Annu Rev Biomed Eng 2014;16:295–320. 
    Crossref | PubMed
  17. Mir LM, Orlowski S, Belehradek J, Paoletti C. Electrochemotherapy potentiation of antitumor effect of bleomycin by local electric pulses. Eur J Cancer 1991;27:68–72. 
    Crossref | PubMed
  18. Miklavčič D, Mali B, Kos B, et al. Electrochemotherapy: from the drawing board into medical practice. Biomed Eng Online 2014;13:29. 
    Crossref | PubMed
  19. Campana LG, Edhemovic I, Soden D, et al. Electrochemotherapy: emerging applications technical advances, new indications, combined approaches, and multi-institutional collaboration. Eur J Surg Oncol 2019;45:92–102. 
    Crossref | PubMed
  20. Potočnik T, Sachdev S, Polajžer T, et al. Efficient gene transfection by electroporation: in vitro and in silico study of pulse parameters. Appl Sci 2022;12:8237. 
    Crossref
  21. Davalos RV, Mir IL, Rubinsky B. Tissue ablation with irreversible electroporation. Ann Biomed Eng 2005;33:223–31. 
    Crossref | PubMed
  22. Lavee J, Onik G, Mikus P, Rubinsky B. A novel nonthermal energy source for surgical epicardial atrial ablation: irreversible electroporation. Heart Surg Forum 2007;10:e162–7. 
    Crossref | PubMed
  23. Batista Napotnik T, Polajžer T, Miklavčič D. Cell death due to electroporation: a review. Bioelectrochemistry 2021;141:107871. 
    Crossref | PubMed
  24. Peng W, Polajžer T, Yao C, Miklavčič D. Dynamics of cell death due to electroporation using different pulse parameters as revealed by different viability assays. Ann Biomed Eng 2024;52:22–35. 
    Crossref | PubMed
  25. Mercadal B, Beitel-White N, Aycock KN, et al. Dynamics of cell death after conventional IRE and H-FIRE treatments. Ann Biomed Eng 2020;48:1451–62. 
    Crossref | PubMed
  26. Miklavčič D, Verma A, Krahn PRP, et al. Biophysics and electrophysiology of pulsed field ablation in normal and infarcted porcine cardiac ventricular tissue. Sci Rep 2024;14:32063. 
    Crossref | PubMed
  27. Markelc B, Bellard E, Sersa G, et al. Increased permeability of blood vessels after reversible electroporation is facilitated by alterations in endothelial cell-to-cell junctions. J Control Release 2018;276:30–41. 
    Crossref | PubMed
  28. Corovic S, Lackovic I, Sustaric P, et al. Modeling of electric field distribution in tissues during electroporation. Biomed Eng OnLine 2013;12:16. 
    Crossref | PubMed
  29. Kranjc M, Bajd F, Serša I, Miklavčič D. Magnetic resonance electrical impedance tomography for measuring electrical conductivity during electroporation. Physiol Meas 2014;35:985–96. 
    Crossref | PubMed
  30. Jarm T, Cemazar M, Miklavcic D, Sersa G. Antivascular effects of electrochemotherapy: implications in treatment of bleeding metastases. Expert Rev Anticancer Ther 2010;10:729–46. 
    Crossref | PubMed
  31. Hjouj M, Last D, Guez D, et al. MRI study on reversible and irreversible electroporation induced blood brain barrier disruption. PLoS One 2012;7:e42817. 
    Crossref | PubMed
  32. Sharabi S, Last D, Daniels D, et al. The effects of point-source electroporation on the blood–brain barrier and brain vasculature in rats: an MRI and histology study. Bioelectrochemistry 2020;134:107521. 
    Crossref | PubMed
  33. Ivorra A, Al-Sakere B, Rubinsky B, Mir LM. In vivo electrical conductivity measurements during and after tumor electroporation: conductivity changes reflect the treatment outcome. Phys Med Biol 2009;54:5949–63. 
    Crossref | PubMed
  34. Iacopino S, Fabiano G, Malacrida M, Petretta A. Lesion effects in terms of local impedance variations after pulsed-field ablation during pulmonary vein isolation: a case report. Eur Heart J Case Rep 2023;7:ytad187. 
    Crossref | PubMed
  35. Avazzadeh S, Dehkordi MH, Owens P, et al. Establishing electroporation thresholds for targeted cell specific cardiac ablation in a 2D culture model. J Cardiovasc Electrophysiol 2022;33:2050–61. 
    Crossref | PubMed
  36. Casciola M, Keck D, Feaster TK, Blinova K. Human cardiomyocytes are more susceptible to irreversible electroporation by pulsed electric field than human esophageal cells. Physiol Rep 2022;10:e15493. 
    Crossref | PubMed
  37. Kos B, Mattison L, Ramirez D, et al. Determination of lethal electric field threshold for pulsed field ablation in ex vivo perfused porcine and human hearts. Front Cardiovasc Med 2023;10:1160231. 
    Crossref | PubMed
  38. Kawamura I, Reddy VY, Santos-Gallego CG, et al. Electrophysiology, pathology, and imaging of pulsed field ablation of scarred and healthy ventricles in swine. Circ Arrhythm Electrophysiol 2023;16:e011369. 
    Crossref | PubMed
  39. Fallert MA, Mirotznik MS, Downing SW, et al. Myocardial electrical impedance mapping of ischemic sheep hearts and healing aneurysms. Circulation 1993;87:199–207. 
    Crossref | PubMed
  40. Schwartzman D, Chang I, Michele JJ, et al. Electrical impedance properties of normal and chronically infarcted left ventricular myocardium. J Interv Card Electrophysiol 1999;3:213–24. 
    Crossref | PubMed
  41. Fiserova I, Fiser O, Novak M, et al. Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro. Heart Rhythm 2025;22:466–74. 
    Crossref | PubMed
  42. Nies M, Koruth JS, Mlček M, et al. Hemolysis after pulsed field ablation: impact of lesion number and catheter-tissue contact. Circ Arrhythm Electrophysiol 2024;17:e012765. 
    Crossref | PubMed
  43. Stojadinović P, Ventrella N, Alfredová H, et al. Prediction of major intravascular hemolysis during pulsed electric field ablation of atrial fibrillation using a pentaspline catheter. J Cardiovasc Electrophysiol 2024;35:2405–10. 
    Crossref | PubMed
  44. Gunawardene MA, Schaeffer BN, Jularic M, et al. Coronary spasm during pulsed field ablation of the mitral isthmus line. JACC Clin Electrophysiol 2021;7:1618–20. 
    Crossref | PubMed
  45. Reddy VY, Petru J, Funasako M, et al. Coronary arterial spasm during pulsed field ablation to treat atrial fibrillation. Circulation 2022;146:1808–19. 
    Crossref | PubMed
  46. Howard B, Haines DE, Verma A, et al. Characterization of phrenic nerve response to pulsed field ablation. Circ Arrhythm Electrophysiol 2022;15:e010127. 
    Crossref | PubMed
  47. Livia C, Sugrue A, Witt T, et al. Elimination of Purkinje fibers by electroporation reduces ventricular fibrillation vulnerability. J Am Heart Assoc 2018;7:e009070. 
    Crossref | PubMed
  48. van Zyl M, Ladejobi AO, Tri JA, et al. Reversible atrioventricular conduction impairment following bipolar nanosecond electroporation of the interventricular septum. JACC Clin Electrophysiol 2021;7:255–7. 
    Crossref | PubMed
  49. Shen C, Bai R, Jia Z, et al. Unexpected transient atrioventricular block and slow junctional rhythm using pulsed field ablation for slow pathway modification: excited or cautious for ablators. Heart Rhythm 2025;22:1682–92. 
    Crossref | PubMed
  50. Katritsis DG. Pulsed field ablation for atrioventricular junctional reentry: simple is complicated. Heart Rhythm 2025;22:1693–5. 
    Crossref | PubMed
  51. Koop B. Fundamentals of system design for cardiac pulsed field ablation: optimization of safety, efficacy, and usability. Pacing Clin Electrophysiol 2025;pace.15120; epub ahead of press. 
    Crossref | PubMed
  52. Weaver JC, Smith KC, Esser AT, et al. A brief overview of electroporation pulse strength–duration space: a region where additional intracellular effects are expected. Bioelectrochemistry 2012;87:236–43. 
    Crossref | PubMed
  53. Batista Napotnik T, Reberšek M, Vernier PT, et al. Effects of high voltage nanosecond electric pulses on eukaryotic cells (in vitro): a systematic review. Bioelectrochemistry 2016;110:1–12. 
    Crossref | PubMed
  54. Arena CB, Sano MB, Rossmeisl JH, et al. High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction. Biomed Eng Online 2011;10:102. 
    Crossref | PubMed
  55. Reddy VY, Koruth J, Jais P, et al. Ablation of atrial fibrillation with pulsed electric fields: an ultra-rapid, tissue-selective modality for cardiac ablation. JACC Clin Electrophysiol 2018;4:987–95. 
    Crossref | PubMed
  56. Reddy VY, Neuzil P, Koruth JS, et al. Pulsed field ablation for pulmonary vein isolation in atrial fibrillation. J Am Coll Cardiol 2019;74:315–26. 
    Crossref | PubMed
  57. Reddy VY, Peichl P, Anter E, et al. A focal ablation catheter toggling between radiofrequency and pulsed field energy to treat atrial fibrillation. JACC Clin Electrophysiol 2023;9:1786–801. 
    Crossref | PubMed
  58. Anić A, Phlips T, Brešković T, et al. Pulsed field ablation using focal contact force-sensing catheters for treatment of atrial fibrillation: 1-year outcomes of the ECLIPSE AF study. Circ Arrhythm Electrophysiol 2025;18:e012794. 
    Crossref | PubMed
  59. Dermol J, Miklavčič D. Mathematical models describing Chinese hamster ovary cell death due to electroporation in vitro. J Membr Biol 2015;248:865–81. 
    Crossref | PubMed
  60. Aycock KN, Campelo SN, Davalos RV. A comparative modeling study of thermal mitigation strategies in irreversible electroporation treatments. J Heat Transfer 2022;144:031206. 
    Crossref | PubMed
  61. Cvetkoska A, Maček-Lebar A, Trdina P, et al. Muscle contractions and pain sensation accompanying high-frequency electroporation pulses. Sci Rep 2022;12:8019. 
    Crossref | PubMed
  62. Cvetkoska A, Maček-Lebar A, Polajžer T, et al. The effects of interphase and interpulse delays and pulse widths on induced muscle contractions, pain and therapeutic efficacy in electroporation-based therapies. J Cardiovasc Dev Dis 2023;10:490. 
    Crossref | PubMed
  63. Kumru HT, Mattison L, Tarakji KG, et al. Arrhythmogenicity of monophasic and biphasic PFA waveforms in a porcine model. J Cardiovasc Electrophysiol 2024;35:2487–90. 
    Crossref | PubMed
  64. Mahnič-Kalamiza S, Miklavčič D. Scratching the electrode surface: insights into a high-voltage pulsed-field application from in vitro & in silico studies in indifferent fluid. Electrochim Acta 2020;363:137187. 
    Crossref
  65. Van Es R, Groen MHA, Stehouwer M, et al. In vitro analysis of the origin and characteristics of gaseous microemboli during catheter electroporation ablation. J Cardiovasc Electrophysiol 2019;30:2071–9. 
    Crossref | PubMed
  66. Wittkampf FH, van Driel VJ, van Wessel H, et al. Feasibility of electroporation for the creation of pulmonary vein ostial lesions. J Cardiovasc Electrophysiol 2011;22:302–9. 
    Crossref | PubMed
  67. Mahnič-Kalamiza S, Miklavčič D, Lombergar P, et al. Elucidating the mechanisms of microbubble formation in intracardiac pulsed field ablation. Electrochim Acta 2024;497:144550. 
    Crossref
  68. Verma A, Zhong P, Castellvi Q, et al. Thermal profiles for focal pulsed electric field ablation. JACC Clin Electrophysiol 2023;9:1854–63. 
    Crossref | PubMed
  69. Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol 1996;41:2251–69. 
    Crossref | PubMed
  70. Gabriel C, Peyman A, Grant EH. Electrical conductivity of tissue at frequencies below 1 MHz. Phys Med Biol 2009;54:4863–78. 
    Crossref | PubMed
  71. Howard B, Verma A, Tzou WS, et al. Effects of electrode–tissue proximity on cardiac lesion formation using pulsed field ablation. Circ Arrhythm Electrophysiol 2022;15:e011110. 
    Crossref | PubMed
  72. Doshi SK, Flaherty MC, Laughner J, et al. Catheter–tissue contact optimizes pulsed electric field ablation with a large area focal catheter. J Cardiovasc Electrophysiol 2024;35:765–74. 
    Crossref | PubMed
  73. Okumura Y, Watanabe R, Nagashima K, et al. In vivo assessment of catheter–tissue contact using tissue proximity indication and its impact on cardiac lesion formation in pulsed field ablation. Heart Rhythm 2025;22:952–60. 
    Crossref | PubMed
  74. Groen MHA, Fish JM, Loh P, Van Es R. Impact of electrode–tissue proximity on formation of PFA lesions: using multi-electrode impedance measurements. Heart Rhythm O2 2025;6:1412–8. 
    Crossref
  75. Saito Y, Nagashima K, Watanabe R, et al. Clinical importance of tissue proximity indication during pulsed field ablation for atrial fibrillation: insights from initial experience. Heart Rhythm 2025;22:e51–9. 
    Crossref | PubMed
  76. Di Biase L, Marazzato J, Govari A, et al. Pulsed field ablation index-guided ablation for lesion formation: impact of contact force and number of applications in the ventricular model. Circ Arrhythm Electrophysiol 2024;17:e012717. 
    Crossref | PubMed
  77. Nakagawa H, Castellvi Q, Neal R, et al. Effects of contact force on lesion size during pulsed field catheter ablation: histochemical characterization of ventricular lesion boundaries. Circ Arrhythm Electrophysiol 2024;17:e012026. 
    Crossref | PubMed
  78. Mattison L, Verma A, Tarakji KG, et al. Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue. J Cardiovasc Electrophysiol 2023;34:693–9. 
    Crossref | PubMed
  79. Eide AR, Mickelson SK, Eide CL, et al. Engineering Fundamentals and Problem Solving. 8th ed. New York: McGraw Hill, 2023.
  80. Mercadal B, Arena CB, Davalos RV, Ivorra A. Avoiding nerve stimulation in irreversible electroporation: a numerical modeling study. Phys Med Biol 2017;62:8060–79. 
    Crossref | PubMed
  81. Pirc E, Miklavčič D, Uršič K, et al. High frequency and high-voltage asymmetric bipolar pulse generator for electroporation based technologies and therapies. Electronics 2021;10:1203. 
    Crossref
  82. Shiba K. Measurement of high-frequency patient leakage current from a transmitting coil in a transcutaneous energy transmission system for ventricular assist device. IEEE Trans Electromagn Compat 2022;64:1304–12. 
    Crossref
  83. International Electrotechnical Commission. 60601-1:2005: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance. 2005.
  84. International Electrotechnical Commission. 60601-1-2:2014 Medical Electrical Equipment – Part 1–2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances – Requirements and Tests. 2014.
  85. Furgał J, Kuniewski M, Pająk P. Analysis of internal overvoltages in transformer windings during transients in electrical networks. Energies 2020;13:2644. 
    Crossref
  86. Bertacchini C. Cliniporator: medical electroporation of tumors. In: Miklavcic D, ed. Handbook of Electroporation. Cham, Switzerland: Springer International, 2017; 1–36. 
    Crossref
  87. Scherr D, Turagam MK, Maury P, et al. Repeat procedures after pulsed field ablation for atrial fibrillation: MANIFEST-REDO study. Europace 2025;27:euaf012. 
    Crossref | PubMed
  88. Watanabe K, Nies M, Reddy VY, Koruth JS. Lesion morphometry of the pentaspline pulsed field ablation catheter: understanding catheter pose, rotation, and dosing. Circ Arrhythm Electrophysiol 2024;17:e013208. 
    Crossref | PubMed
  89. Wijffels MCEF, Van Oosterhout M, Boersma LVA, et al. Characterization of in vitro and in vivo lesions made by a novel multichannel ablation generator and a circumlinear decapolar ablation catheter. J Cardiovasc Electrophysiol 2009;20:1142–8. 
    Crossref | PubMed
  90. Witt CM, Sugrue A, Padmanabhan D, et al. Intrapulmonary vein ablation without stenosis: a novel balloon-based direct current electroporation approach. J Am Heart Assoc 2018;7:e009575. 
    Crossref | PubMed
  91. Reichlin T, Kueffer T, Badertscher P, et al. Pulsed field or cryoballoon ablation for paroxysmal atrial fibrillation. N Engl J Med 2025;392:1497–507. 
    Crossref | PubMed
  92. Reddy VY, Gerstenfeld EP, Natale A, et al. Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation. N Engl J Med 2023;389:1660–71. 
    Crossref | PubMed